The MFH-500 ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. It is mainly intended for providing ventilation support to a critically ill patient during the life threatening phase and ensuring the going-through of the dangerous period by the patient and smooth treatment of primary diseases for recovery. Also it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation.
Component Details
Display: LED display following parameters
Operating mode and respiratory rate;
--Ventilator control frequency set by medical staff
--Tidal volume of each breath in the patient
--Occurrence of spontaneous breath in the patient and spontaneous breathing frequency
--Expiratory phase and inspiratory phase and actual respiratory rate
Ventilator construction
--Environmental protection material construction, hardly to be broken
--Trolly: fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Techniques soecification
Physical specification |
Display | LED display |
Mode | pneumatically pwoered and electrically controlled system |
Trolley | fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions |
Technical specifications |
Gas requirement | Medical oxygen with a pressure ranging from 280-600kpa |
safety valve respiratory rate oxygen concentration output | <12.5kPa 6-60 times/min <50% |
minute volume | ≥18L |
modes of ventilation | A/C, IPPV, SIPPV, IMV, SIMV, PEET, MANUAL, SIGH |
PEEP | 0.1-1.0kPa |
Ptr | -4-10hPa |
IMV frequency | 1-12bpm |
I/E Ratio | 1:1.5, 1:2.0; 1:2.5; 1:3.0 |
Tidal volume | Not less than 50-1200ml |
Inspiratory plateau | 0-1s |
Max safety pressure | ≤6.0 KPa |
Sigh | one deep breath per 80 controlled respirations, the insporation time is 1.5 times of the setting point |
Pressure limit range | 0-6.0kPa |
Airway pressure alarm | Lower: 0.9kPa-5.4kPa; Upper: 0.5kPa |
Audible and visual and with yellow and red color indicating | ±0.2kPa |
Ventilation colume alarm | ≥18L/min |
Power supply alarm | AC/DC power supply are after failing to send our alarm immediately Alarm time: keep >120s |
The airway pressure continues to be higher than 15hPa±1hPa for 15s±1s, the the machine will raise an audible alarm |
Operating condition |
Ambient temperature | 10-40ºC |
Relative humidity | no higher than 80% |
Atmospheric pressure | 860kPa-1060kPa |
Power requirement | 100-120 Vac, 50/60Hz |
Attention: the AC power supply used for the anesthesia machine must be well grounded. Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO9918:1993 |
Storage |
Ambient temperature | -15ºC-+50ºC |
Relative humidity | no higher than 95% |
Atmospheric pressure | 86kPa-106kPa |
| It should be stored in a room without corrosive gas and well ventilated |
Package |
Packagin box | comply with the requirement of GB/T 15464 |
| between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation |
| moisture protection and rain protection to ensure that product is protected from natural damage |
Safety & Alarm |
Gas source pressure | immediate alarm; "high level alarm" indicator flickers raising an audible alarm |
Airway pressure alarm | the airway pressure exceeds 15hPa. the max deplay is no longer than 17s. the "high level alarm" indicator flickers raising an audible alarm |
Tidal volume alarm | Lower than 50ml, the alarm will be raised after one breathing cycle |
Ventilation volume alarm | Lower: 0-12L/min; Up: 18L/min |
Battery voltage alarm | it alarms when the battery voltage is lower than 10.5V |
Power on self test | self test before working, assure it's working normally |
Standard configurations |
QTY | NAME |
1 set | main unit |
1 set | oxygen tube |
2 sets | silicone mask |
1 pc | power line |
2 pcs | threaded pipe |
2 pcs | standard pipe |
1 set | oxygen reducer |
1 set | humidifier |
1 pc | battery |
1 set | flow sensor |
2 pcs | a leather bag (blue) |
1 set | tools with the machine |
1 set | user manual (english version) |