Ventilator, Blower, Ventilation Fan manufacturer / supplier in China, offering ICU Ventilator Mfh-500 with CE Certificate, Mfj-Yx280as Popular Stainless Steel Autoclaves (time-controlled type, when the control + anti-dry type), Temp Controlled Multifunctional Prp. Ppp. Fat Centrifuge Mfl5-G and so on.

(/ )

Supplier Homepage Product Ventilator ICU Ventilator Mfh-500 with CE Certificate

ICU Ventilator Mfh-500 with CE Certificate

FOB Price: Get Latest Price
Min. Order: 1 Piece
Production Capacity: 3000PCS/Year
Payment Terms: L/C, T/T

Request a custom order and have something just for you!

Send Customized Request
Basic Info
  • Model NO.: MFH-500
  • Ventilation Capacity Per Minute: ≥18L
  • Ventilator Conformance: ≤3
  • Trademark: MEDFAR
  • Origin: Shandong
  • Tidal Volume Adjustment: 50~1500ml
  • Oxygen Concentration: 21%~100%
  • Respiratory Rate: 1~99/M
  • Specification: CE & ISO
Product Description
The MFH-500 ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. It is mainly intended for providing ventilation support to a critically ill patient during the life threatening phase and ensuring the going-through of the dangerous period by the patient and smooth treatment of primary diseases for recovery. Also it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation.

Component Details

Display: LED display following parameters
Operating mode and respiratory rate;
--Ventilator control frequency set by medical staff
--Tidal volume of each breath in the patient
--Occurrence of spontaneous breath in the patient and spontaneous breathing frequency
--Expiratory phase and inspiratory phase and actual respiratory rate

Ventilator construction
--Environmental protection material construction, hardly to be broken
--Trolly: fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions

Techniques soecification
Physical specification
DisplayLED display
Modepneumatically pwoered and electrically controlled system
Trolleyfitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Technical specifications
Gas requirementMedical oxygen with a pressure ranging from 280-600kpa
safety valve
respiratory rate
oxygen concentration output
<12.5kPa
6-60 times/min
<50%
minute volume≥18L
modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET, MANUAL, SIGH
PEEP0.1-1.0kPa
Ptr-4-10hPa
IMV frequency1-12bpm
I/E Ratio1:1.5, 1:2.0; 1:2.5; 1:3.0
Tidal volumeNot less than 50-1200ml
Inspiratory plateau0-1s
Max safety pressure≤6.0 KPa
Sighone deep  breath per 80 controlled respirations, the insporation time is 1.5 times of the setting point
Pressure limit range0-6.0kPa
Airway pressure alarmLower: 0.9kPa-5.4kPa; Upper: 0.5kPa
Audible and visual and with yellow and red color indicating±0.2kPa
Ventilation colume alarm≥18L/min
Power supply alarmAC/DC power supply are after failing to send our alarm immediately
Alarm time: keep >120s
The airway pressure continues to be higher than 15hPa±1hPa for 15s±1s, the the machine will raise an audible alarm
Operating condition
Ambient temperature10-40ºC
Relative humidityno higher than 80%
Atmospheric pressure860kPa-1060kPa
Power requirement100-120 Vac, 50/60Hz
Attention: the AC power supply used for the anesthesia machine must be well grounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO9918:1993
Storage
Ambient temperature-15ºC-+50ºC
Relative humidityno higher than 95%
Atmospheric pressure86kPa-106kPa

It should be stored in a room without corrosive gas and well ventilated
Package
Packagin boxcomply with the requirement of GB/T 15464

between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation

moisture protection and rain protection to ensure that product is protected from natural damage
Safety & Alarm
Gas source pressureimmediate alarm; "high level alarm" indicator flickers raising an audible alarm
Airway pressure alarmthe airway pressure exceeds 15hPa. the max deplay is no longer than 17s. the "high level alarm"  indicator flickers raising an audible alarm
Tidal volume alarmLower than 50ml, the alarm will be raised after one breathing cycle
Ventilation volume alarmLower: 0-12L/min; Up: 18L/min
Battery voltage alarmit alarms when the battery voltage is lower than 10.5V
Power on self testself test before working, assure it's working normally
Standard configurations
QTYNAME
1 setmain unit
1 setoxygen tube
2 setssilicone mask
1 pcpower line
2 pcsthreaded pipe
2 pcsstandard pipe
1 setoxygen reducer
1 sethumidifier
1 pcbattery
1 setflow sensor
2 pcsa leather bag (blue)
1 settools with the machine
1 setuser manual (english version)